Contact an Account ManagerCharacterization of ageing products of ester-based synthetic lubricants
Summary
Type: HPLC
Application #: 117590
Catalog#: 720001.40 - MN HPLC Column, Nucleosil C8, 100A, 5um, 4x125mm
Detection: mass spectrometry
Conditions: Eluents: water, methanol
Gradient: water-methanol gradient (80% methanol to 98% methanol in 15 min with a constant amount of 2% ammonium acetate buffer, prepared as a 5 mM solution of ammonium acetate in water) followed by isocratic elution for 30 min
Flow rate: 200 μl/min
Authors: Kohler M Heeb N V
Source: Journal of Chromatography A 926 (2001) 161-165